Appendix D. Performance Test
Test II. Output Power
b Press S
INGLE
If the curve on the display is not clear, you can change resolution by
using the S
key and the Vernier knob.
PAN
5 From the displayed measurements check and note the value for
the Mean Wavelength in the test record.
Test II. Output Power
1 Set up and switch on the equipment:
a Connect the equipment as shown in Figure D-2.
HP E5574 Optical Loss Analyzer
DUT
Head A
Head B
Input
Output
Optical Interface
#21: HP 81000AI
#22: HP 81000SI
SM-Fiber
#21: HP 81101AC
#22: HP 81102SC

Output Power Test Setup

b Ensure that the instruments have warmed up and that the
laser has settled.
2 Zero the HP 8153A:
a Mount the protection cap on the head in order to keep the
light out.
b Press M
ENU
c
Press Z
ERO
199
S
.
WEEP
HP 8153A
Standard Lightwave Multimeter
Standard
Optical Head
Optical Adapter
#20: HP 81000FA
#21, #22: HP 81000AA
to change to Measure Mode.
and wait while zeroing.
Standard Optical Head
Interface Module
HP 81533B