5 Check to see if the analyzer is phase locking to the external CW signal:
6 On the external source, press CW, 50, MHz/µ.
7 On the analyzer, press 50, M/µ.
8 Repeat
In Case of
Difficulty
1 Be sure the external source power is set within 0 to –25 dBm.
2 Make sure the analyzer's "Ext Source Auto" feature is selected. In addition,
3 Verify that all connections are tight.
Required
Equipment
Performance Tests

3. Test Port Output Power Accuracy

• If the analyzer displays any phase lock error messages, write "unlock" in the
"Performance Test Record" for the set CW signal.
• If the analyzer does not display any phase lock error messages, write "lock"
in the "Performance Test Record" for the set CW signal.
Step 5
through
listed in the "Performance Test Record." For analyzers with Option 006, change
the power sensor to the Agilent 8481A for frequencies above 4 GHz.
verify that the analyzer is set to measure its input channel R.
3. Test Port Output Power Accuracy
Perform this test to confirm the accuracy of the Agilent 8702D source output
power.
Equipment Required for 50Ω Analyzers
Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 436A/437B/438A
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8482A
Adapter, APC-7 to Type-N (f) . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 11524A
Additional Equipment Required for Analyzers with Option 006
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8481A
Equipment Required for 75Ω Analyzers
Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 436A/437B/438A
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8483A Option H03
4-6
Step 7
for the other external source CW frequencies