Analog Test Failure Troubleshooting in the Internal Test

When "Analog Test: Fail" is displayed in the internal test, replace the faulty assembly as
shown in Table 4-1. The contents of the analog test is as follows;
Table 4-1
Troubleshooting Information for Analog Test Failure
Test
First Failed Test
No.
1
DC_A1_P9V
+9V on A1
2
DC_A1_N9V
-9V on A1
3
DC_A1_P5VRF1
+5V for RF1 on A1
4
DC_A1_P5VRF2
+5V for RF2 on A1
5
DC_A1_P5V
+5V on A1
6
DC_A3_P5VPLL
+5V for PLL on A3
7
DC_A3_P10V
-10V on A3
8
DC_A3_N5V
-5V on A3
9
DC_A3_N10VPLL
-10V on A3
10
DC_A3_OUTDET
Detect Level of RFOUT AGC on A3
11
DC_A3_LODET
Detect Level of Local AGC on A3
12
DC_A4_P10V
+10.5V on A4
13
DC_A4_P5V
+5V on A4
14
DC_A4_P8VA
+8V for CH1 on A4
15
DC_A4_P8VB
+8V for CH2 on A4
16
DC_A4_N10_5V
-10.5V on A4
17
DC_A4_N5V
-5V on A4
18
DC_A4_LODET
Detect Level of Local AGC on A4
19
A2_STAT
Status of A2
20
DC_A2_P50V
+50V on A2
21
DC_A2_N50V
-50V on A2
22
DC_A2_P5VREF
+5V for Reference on A2
23
FREQ_A1_VCO1
VCO Output for RF1 on A1
24
FREQ_A1_REF1
PLL Reference for RF1 on A1
25
FREQ_A1_VCO2
VCO Output for RF2 on A1
26
FREQ_A1_REF2
PLL Reference for RF2 on A1
Chapter 4
Test Point
A1 Ref
OSC
†††
†††
†††
†††
†††
†††
†††
†††
†††
Troubleshooting Using Internal Test
Probable Faulty Assembly
A3
A4
A2 DC
Source
Receiver
Bias(001)
†††
†††
†††
†††
†††
†††
†††
†††
†††
†††
†††
†††
††
†††
†††
†††
†††
†††
Troubleshooting
A5 Freq
A6 Test
Ref(1D5)
Head
173