Adjustments and Correction Constants

Analog Bus Correction Constants (Test 44)

Analog Bus Correction Constants (Test 44)
This procedure calibrates the analog bus by using three reference voltages (ground,
+0.37 volts and +2.5 volts), then stores the calibration data as correction constants in
EEPROMs.
Analyzer warmup time: 30 minutes.
1. Make sure the A7 Switch is in the alter position (ALT).
2. Press the following:
PRESET: FACTORY
Preset
SERVICE MENU
System
EXECUTE TEST
44
x1
3. Observe the analyzer for the results of the adjustment routine:
• If the analyzer displays ABUS Cor DONE, you have completed this procedure.
• If the analyzer displays ABUS Cor FAIL, repeat the test.
4. If no more correction constant routines are going to be performed, return the A7 Switch
to the NRM position and perform the
In Case of Difficulty
• Refer to
Chapter 6 , "Digital Control Troubleshooting,"
Technologies sales or service office.
3-8
Preset
TESTS
YES
"EEPROM Backup Disk Procedure" on page
8719ET/20ET/22ET
8719ES/20ES/22ES
RF Network Analyzers
or contact the nearest Agilent
Service Guide
3-33.