Installation

Packing the instrument for transportation

Packing the
instrument for
transportation
Refitting the
transport lock
30
Thermo Scientific KingFisher Duo User Manual
Figure 3–17. KingFisher Duo 6-pin head for KingFisher Flex 24 deep well
plate fitted
This section describes how to refit the transport lock when packing the
instrument for transportation or service.
To refit the transport lock:
1. To fit the transport lock, choose the Transportation_lock protocol
under the Maintenance menu by using the Up and Down arrow
keys.
2. Then press the
START
3. Switch off the power.
4. Refit the transport lock into place by screwing the finger screw with
spring suspension clockwise (Figure 3–18). There is only one
possible position for the finger screw to lock once the processing
head is correctly located after Steps 1–2.
key.