7.7 Auto Peak Adjust

For optimum performance it is important that the analyte wavelengths are correctly aligned in the
centre of the sub array measurement window. The iCAP 7000 Series Spectrometer will
automatically check a reference line each time the plasma is ignited to maintain wavelength
accuracy.
Auto Peak only needs to be run whenever a new wavelength is used for analysis. The
autopeak may also need to be run if the instrument has been switched off for an extended period
of time.
To carry out an Auto Peak, allow the plasma to stabilize for 10 minutes and in the Acquisition
Parameters of the Methods Parameters of your LabBook Qtegra select Perform Auto Peak.
This document is part numbered 849940090071
Version 1.02
Page 37 of 72