Adjust To Abovevial Pusher Position - Thermo Scientific VF-F20-A Operating Manual

Integral fraction collector ft
Table of Contents

Advertisement

6.8.1

Adjust to AboveVial Pusher Position

Preparations
Follow these steps
Integral Fraction Collector FT (VF-F20-A) Operating Manual
No.
Description
1
Lock pin
2
Pusher cover
1. Open the left front
2. Press the service button on keypad to move the needle unit into
service position.
NOTICE To avoid personal injury, do not put hands into the
compartment when the needle unit is moving.
1. Raise the lock pin upwards by hand. The pusher will be moved up
accordingly.
Figure 52: Moving the pusher up by hand
No.
Description
1
Lock tab
No.
Description
3
Pusher
4
Needle
door.
No.
Description
2
U-shape slot
6
Operation
Page 133