12.2 Troubleshooting Flow Chart

12-8
Nothing is displayed.
Is the power switch on?
Does the switch work properly?
YES
Is the power supply
connected properly?
YES
Is the power supplied properly?
YES
Is the fuse blown?
YES
Replace the fuse (See section 13.2).
YES
Does the fuse stay intact
when the power is turned
on again?
YES
Contact Omega.
NO
Turn on the power properly
(See section 3.2).
NO
Connect properly (See section 2.7).
Supply the power at the specified
NO
voltage and frequency.
(See section 14.7)
NO
Contact Omega.
NO
Contact Omega.