Omega RD-MV100 Series User Manual page 270

Hide thumbs Also See for RD-MV100 Series:
Table of Contents

Advertisement

14.3 Data Storage Specifications
14-6
Internal memory capacity
Data Type
Display data only
Display data and event data
Event data only
Maximum number of data points per channel that can be stored
Data Type
Display data only
Display data and event data
Event data only
Capacity of Internal Memory
1.2 MB
Display data : 0.9 MB
Event data : 0.3 MB
1.2 MB
Maximum Number of Data Points Per Channel
1,200,000 bytes/(number of measurement channels × 4 +
number of computation channels × 8)
Except, the maximum number of data points is 100,000
Display data
900,000 bytes/(number of measurement channels × 4 +
number of computation channels × 8)
Except, the maximum number of data points is 75,000
Event data
300,000 bytes/(number of measurement channels × 2 +
number of computation channels × 4)
Except, the maximum number of data points is 30,000
1,200,000 bytes/(number of measurement channels × 2 +
number of computation channels × 4)
Except, the maximum number of data points is 120,000
This logic is explained in more detail below :
1) When acquiring display data only
If we assume that the number of measuring channels is
20, the number of computing channels is 10, and the
display rate is 30 min/div (60 sec sampling interval),
then :
Number of data per channel = 1,200,000 bytes/(20 × 4
bytes + 10 × 8 bytes) = 7,500 data
* Maximum number of data points is 100,000.
Sampling length per file = 7,500 × 60 sec = 450,000 sec
= approx. 5 days
2) When acquiring event data only
If we assume that the number of measuring channels is
20, the number of computing channels is 10, and the
sampling interval is 1 sec, then :
Number of data per channel = 1,200,000 bytes/(20 × 2
bytes + 10 × 4 bytes) = 15,000 data
* Maximum number of data points is 120,000.
Sampling length = 15,000 × 1 sec = 15,000 sec =
approx. 4 hours
3) When acquiring both display data and event data
The sampling length is calculated by defining the
capacity for display data as 900,000 bytes and the
capacity for event data as 300,000 bytes. The method
of calculation is the same as shown above.
Except, the maximum number of data points is 75,000
for display data and 30,000 for event data.
*
*

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rd-mv102Rd-mv104Rd-mv106Rd-mv112

Table of Contents