Omega RD-MV100 Series User Manual page 52

Hide thumbs Also See for RD-MV100 Series:
Table of Contents

Advertisement

2.3 Input Signal Wiring
DC Voltage and DI (Contact) Input
+
DC voltage input
Resistance Temperature Detector Input
A
B
b
Leadwire resistance :
The resistance of the three wires
should be approximately equal.
2-6
• Wiring diagram
+
b
A
B
10 Ω max./wire
Thermocouple Input
Extension leadwire
+
DC Current Input
+
+
DC current input
Shunt resistor
NOTE : For a 4 to 20 mA input, use a
shunt resistor of 250 Ω ±0.1%.