Omega RD-MV100 Series User Manual page 60

Hide thumbs Also See for RD-MV100 Series:
Table of Contents

Advertisement

2.7 Power Supply Wiring
12 VDC Power Supply Model
2-14
Using AC adapter
Follow the warnings below to avoid electric shock or damaging the instrument.
• Connect the power wires after checking that the power supply is turned off to
prevent electric shock.
• To prevent electric shock or fire, always use the power cable supplied by
Omega.
• Before connecting the power cord, check that the voltage on the supply side
matches with the voltage rating of this recorder.
• When not using the instrument for a long time, unplug the power cord of the AC
adapter.
• Only use the AC adapter from Omega (Part No.: B9230ZW).
• Do not put objects on top of the AC adapter or the power cord. Also, do not let
heat generating objects come in contact with them.
• When unplugging the power cord, do not pull on the power cord. Always hold
the plug. If the power cord becomes damaged, contact your nearest
representative listed on the back cover of this manual.
Connecting Procedure
1. Check that the power switch is turned OFF.
2. Connect the AC adapter to the AC adapter jack on RD-MV100.
Install the clamp filter (standard accessory) near RD-MV100 of power cord.
3. Connect the power cord plug that came with the AC adapter to the power supply
connector of the AC adapter.
4. Connect the plug on the other end of the power cord to a power outlet meeting the
following specifications.
The power outlet should be a three-pole type with a protective grounding terminal.
Rated power supply voltage:
Operating supply voltage range:
Rated supply voltage frequency:
Maximum power consumption:
AC adapter rated output voltage:
AC adapter maximum rated output current:
Functional grounding
When using the AC adapter, noise may be reduced if the functional ground is connected
to the earth GND. Use the functional ground terminal as necessary.
WARNING
100 to 240 VAC
90 to 250 VAC
50/60 Hz
90 VA
12 VDC
2.6 A

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rd-mv102Rd-mv104Rd-mv106Rd-mv112

Table of Contents