14.6 Specifications of Optional Functions
3 terminal Isolated RTD Input (/N2) :
12 VDC Power Supply:
Remote Control (/R1) :
14-12
A, B, b terminals are of isolated input type.
* Can be specified only for RD-MV106 and RD-MV112.
A,B,b terminals of RD-MV102 and RD-MV104 are isolated as standard.
Powered by 12 VDC
For related specifications, see "Power Supply," "Normal Operating Conditions," "Effects
of Operating Conditions," and "Other Specifications" in section 14.7.
Rated power supply :
12 VDC
Allowable power supply voltage range :
10 to 18 VDC
Rated power consumption :
30 VA
This option allows the following functions to be controlled remotely by a contact input (up
to eight contact inputs) :
• Start/stop of memory (rising and falling edge)
• Trigger for event data acquisition (trigger, 250 ms or longer)
• Time adjustment (adjusting the time to a preset time upon contact signal, trigger, 250
ms or longer)
Time of Trigger-On
Cut off reading of less than one minute.
hh:00:00 to hh:01:59
e.g. 10:00:50 is corrected as 10:00:00
Round up reading of less than one minute.
hh:58:00 to hh:59:59
e.g. 10:59:50 is corrected as 11:00:00
hh:02:00 to hh:57:59
No process is to be performed.
• Start/stop of computation (rising and falling edge)
• Reset of computation data (trigger, 250 ms or longer)
• Manual sample (trigger, 250 ms or longer)
• Message display (Up to 8 different messages can be set, trigger, 250 ms or longer)
• Load of setting parameters (Up to 3 settings can be set, trigger, 250 ms or longer)
• Alarm acknowledgement (trigger, 250 ms or longer)