8.3
Function to Acquire Other Data
Manual Sampled Data
• When the manual sample key is pressed, the instantaneous values of all channels
(excluding the measurement channels that are set to [Skip] and the computation
channels that are turned Off) are acquired.
• Up to 50 data sets can be stored in the internal memory. When this number is
exceeded, data are overwritten from the oldest data.
Note
TLOG Data (Option)
• TLOG data are acquired at each time interval set by timers.
• Up to 400 data sets can be stored in the internal memory. When this number is
exceeded, data are overwritten from the oldest data.
Note
Report Data (Option)
• Report data are acquired at each time interval (one hour for hourly report, one day for
daily report, etc).
• Up to 40 data sets can be stored in the internal memory. When this number is
exceeded, data are overwritten from the oldest data.
For "hourly only", this constitutes 40 minutes of report data. For "daily+monthly", this
constitutes 39 daily reports and one monthly report or 38 daily reports and two
monthly reports.
Note
The number of manual sampled data in the internal memory can be confirmed with the
memory summary. ⇒ "Section 4.5"
Up to 16 TLOG files (number of start and stop) can be stored to the internal memory. When
the number of files in the internal memory exceeds 16, TLOG data are overwritten even if the
number of data sets is less than 400.
The number of TLOG data sets in the internal memory can be confirmed with the memory
summary. ⇒ "Section 4.5"
The number of report data sets in the internal memory can be confirmed with the memory
summary. ⇒ "Section 4.5"