Flow Unit Select

Smart Manometers that use differential pressure sensors can be
programmed to read out flow measurement units such as CFM or
L/min. The primary element must be a differential pressure -
square root type device such as a pitot tube, orifice plate or venturi.
The flow constant and flow units description are programmed into
the manometer using the same keystrokes used in the User Unit
Select programming. At step 3 choose "FLOW UNIT SELECT"
instead of "USER UNIT SELECT".
Calculate the Flow constant from the following equation:
F
= Q ÷ DP
c
where: F
= Flow constant
c
Q = Flow rate
DP = Differential pressure corresponding to Q
Example: If the DP is 25 units when the flow rate is 10,000 units,
then the Flow constant is 2,000.

Damp Rate Select

Adjustable exponential type damping is available to steady the
display when measuring pulsating pressure or flow. The Smart
Manometer has a range of damping rates; 0.1, 0.2, 0.5, 1, 2, 5, 10,
or 25 seconds. Exponential damping shows approximately 70% of
a step change in pressure upon the next display update. When set
arrow
arrow key. This will toggle the bottom line from
½
(from flow element calculation sheet)
Page 8 of 28
Top line reads "USER UNIT
NAME".
Bottom line reads "FT H2O".
Last letter "O" is blinking.
Top line reads "PROGRAM
MODE".
Bottom line reads "UNITS
SELECT".
Bottom line reads "EXIT".
Manometer returns to Measure
Mode. Units Display shows "FT
H2O".
Step 5

Advertisement

Table of Contents
loading

Table of Contents