8.6 Acquiring Event Data
During [Trigger] Mode or [Rotate] Mode
8-14
Start data acquisition
1.
Press the START key to enter the trigger wait state. The icon in the status
display section will change accordingly. A bar indicating the pre-trigger data in
the status display section is displayed in orange.
2.
When the trigger occurs, data acquisition starts.
Providing the key trigger that starts the data acquisition to the internal memory
This operation is carried out in the trigger wait state. For all other cases, this operation
produces no effect. For details related to triggers other than the key trigger, see section
sections 8.2 or 8.11.
1.
Press the FUNC key to display the soft key menu.
2.
Press the [Trigger] soft key to start the data acquisition of the event data to the
internal memory.
Stop data acquisition
Writing stops upon reaching the specified time (data length). In the status display
section, the block to which data were stored turns green. See section 4.2.
Note
During the [Trigger] mode, [Full] is displayed when data acquisition to all blocks is complete.
When [Full] is displayed, event data are no longer acquired even if the trigger condition is
met. In this case, save the data in the internal memory to the external storage medium.
Terminate data acquisition with a key operation (in sync with the termination of
the display data acquisition when acquiring the display data and the event data)
1.
Press the STOP key. A confirmation window opens.
On models with the computation option, the confirmation window displays the
following choices : [Mem + Math] (terminate acquisition and all computations),
[Memory] (terminate acquisition), and [Cancel].
2.
Select [Yes] ([Mem + Math] or [Memory] for models with the computation
function) using the arrow keys and press the DISP/ENTER key to stop the data
acquisition. The icon in the status display section changes to a stop icon.
Note
One file is created for one set of start and stop operations. The maximum number of files that
can be stored in the internal memory is the specified number of blocks.
When a power disruption occurs the file is closed. When the power resumes, data are written
to a new file.
Memory write operation can also be started or stopped via remote control or communication
commands.
The START/STOP operation also starts/stops the report function.
When the computation is stopped, it is started with the START key.