11.9 Setting the Timer (Basic Setting Mode)

Procedure
11-18
Up to three timers can be specified. The timers specified here are used to determine the
interval for the TLOG computation. In addition, all the data of measurement and
computation channels can be saved at the interval determined by the timers.
• These procedures are carried out in the basic setting mode.
• Use the arrow keys to move the cursor (blue) to the appropriate parameter.
• The possible selections are displayed at the bottom of the screen. Select using the
soft keys.
• The parameter boxes that are changed are displayed in yellow. By confirming the
new settings, they change back to white.
Press
MENU
.
Press
for 3 seconds. The menu screen for the basic setting mode appears.
FUNC
Twice press the soft key
#9
Press the soft key
To display the setting screen press the soft key
The numbers in front of the operating procedure shown below correspond to the
numbers in the above figure.
Setting the relative timer
1.
Timer Number
Select the desired timer number.
2.
Mode
Select [Relative]. [Interval], [Reset], and [Action] are displayed.
3.
Interval
Set the time to the next timeout.
Pressing the [Input] soft key displays a window used to enter the time. Enter the
time (00:01 to 24:00) and press the DISP/ENTER key. For the procedures
related to entering numerical values, see "Entering Numbers" on page 3-21.
4.
Reset
[On] : Resets the TLOG value when the timer expires.
[Off] : Does not reset the TLOG value when the timer expires.
5.
Action
[DataSave] : Stores the instantaneous values of all measurement/computation
channels to the internal memory when the time expires.
[Off] :
Does not store the measured/computed data.
Next 1/3
.
.