11.1 Overview of the Computation Function
Unit Handling
Order of Precedence of the Operators
Displaying the Computed Result
11-2
Communication input data (C01 to C12)
Data that have been specified through the communication interface can be used. Enter
the data as C01 to C12 in the equations. For the procedure used to set the data, see the
"RD-MV100 Communication Interface User's Manual" (IM RD-MV100-17E).
Range of numerical values (Maximum number of significant digits is 5) :
–9.9999E + 29 to –1.0000E – 30, 0, 1.0000E – 30 to 9.9999E + 29
Conditions of the remote control terminals (D01 to D08)
The conditions of the remote input signal can be assigned to "1" and "0," and used in the
equations. Enter the data as D01 to D08 (the number following the letter D is the remote
terminal number) in the equations.
The correlation between the conditions of the remote input signal and the value "1" and
"0" are shown below.
Type of
the Remote
Input Signal
Status
Contact
close
open
Open collector
Voltage level is Lo at the remote terminal
Voltage level is Hi at the remote terminal
The unit corresponding to the measured/computed data in the equation is not
compensated. In computations, measured and computed data are handled as values
without units. For example, if the measured data from channel 01 is 20 mV and the
measured data from channel 02 is 20 V, the computed result of 01 + 02 is 40.
The order of precedence of the operators in the equation is as follows. The operators
are placed in order from the highest precedence.
Type
Functions
Power
Logical negation
Multiplication and division
Addition and subtraction
Greater than/less than
Equal/not equal
Logical AND
Logical OR, exclusive OR
The computed data of computation channels can be displayed in each operation screen.
Displaying the waveform and bar graph
The data are displayed in the range defined by the upper and lower limits of the span.
For the settings related to displaying the computed result, see sections below.
Set computation channels : See section 11.4.
Assign channels to groups : See section 7.6.
Display tags : See sections 7.1 and 7.2.
Set the channel display color : See section 7.8.
Use zone display : See section 7.9.
Set the number of scale divisions : See section 7.10.
Operator
(higher precedence)
ABS( ), SQR( ), LOG( ), EXP( ), TLOG.MAX( ),
TLOG.MIN(), TLOG.P-P( ), TLOG.SUM( ), TLOG.AVE( )
NOT
, /
+, –
.GT., .LT., .GE., .LE.
.EQ., .NE.
AND
OR, XOR
(lower precedence)
"1" or "0"
1
0
1
0

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rd-mv102Rd-mv104Rd-mv106Rd-mv112