Using The Report Function - Omega RD-MV100 Series User Manual

Hide thumbs Also See for RD-MV100 Series:
Table of Contents

Advertisement

11.12 Using the Report Function

Starting/Stopping the Report Function
The report start/stop is in sync with the data acquisition start/stop. The report can also
be started or stopped via remote control or communication commands. For the remote
control function, see section 10.9. For the communication function, see RD-MV100
Communication Interface User's Manual.
Procedure
Starting the report (also starts the data acquisition to the internal memory)
1.
Stopping the report (also stops the data acquisition to the internal memory)
1.
2.
Note
Clearing the Report Data in the Internal Memory
These procedures are carried out in the setting mode. The report data are cleared along
with the display data, for example. For the procedures, see section 9.7.
Press the START key. The icon that indicates the data acquisition to the
internal memory turns to a start icon and the computation icon appears. For
details related to the status display section, see section 4.2.
The report function starts. When the time to create the report arrives, the report
data are written to the internal memory.
Press the STOP key. A confirmation window opens.
Use the arrow keys to select [Mem + Math] (stop the data acquisition to the
internal memory and computation) or [Mem] (stop the data acquisition), and
press the DISP/ENTER key.
The icon that indicates the data acquisition to the internal memory turns to a
stop icon and the computation icon disappears.
The report stops.
When the report function is stopped, the report file is closed. When the report is restarted,
data are written to a new file.
11
11-25

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rd-mv102Rd-mv104Rd-mv106Rd-mv112

Table of Contents