14.7 General Specifications

Construction
Installing:
Dimensions :
Weight :
Standard Performance
Measuring and Recording Accuracy :
Input
DC voltage
TC
(Excluding the
reference
junction
compensation
accuracy)
RTD
Standing on its front feet, the recorder can be inclined backwards up to
30 degrees from a horizontal plane.
approx. 152(W) × 225(H) × 240(D) mm
RD-MV102 : approx. 3.7 kg
RD-MV104 : approx. 3.7 kg
RD-MV106 : approx. 3.8 kg
RD-MV112 : approx. 3.8 kg
The following specifications apply to operation of the recorder under
standard operation conditions :
23 ± 2°C
Temperature :
55% ± 10% RH
Humidity :
Power supply voltage :
Power supply frequency : 50/60 Hz ± 1%
Warm-up time :
At least 30 minutes.
Other ambient conditions such as vibration should not adversely affect
recorder operation.
Range
Measurement Accuracy (Digital Display)
20 mV
60 mV
200 mV
± (0.1% of rdg + 2 digits)
2 V
6 V
20 V
± (0.15% of rdg + 1°C)
R
However,
± 3.7°C at 0 to 100°C,
R, S :
S
± 1.5°C at 100 to 300°C
± 2°C at 400 to 600°C
B :
B
(Accuracy at less than 400°C is not guaranteed.)
± (0.15% of rdg + 0.7°C)
K
However, ± (0.15% of rdg + 1°C) at –200 to –100°C
± (0.15% of rdg + 0.5°C)
E
± (0.15% of rdg + 0.5°C)
J
However, ± (0.15% of rdg + 0.7°C) at –200 to –100°C
T
± (0.15% of rdg + 0.7°C)
N
± (0.15% of rdg + 1°C)
W
± (0.15% of rdg + 0.5°C)
L
However, ± (0.15% of rdg + 0.7°C) at –200 to –100°C
U
Pt100
± (0.15% of rdg + 0.3°C)
JPt100
90 to 132 or 180 to 250 VAC
Max. Resolution of Digital Display
10 µV
10 µV
100 µV
1 mV
1 mV
10 mV
0.1°C