6.1 Releasing the Alarm Indication and Output Relay (Option)
6-2
Non -hold
Activated
Alarm
Released
Red icon
Indication
No icon
• Alarm indication in the trend display
Alarm type
Indicator Hold/Non-Hold
*
Non-hold
Hold
* When the alarm indication is set to non-hold, alarm ACK does not work.
• Alarm indication in the digital display
Measured value
Alarm mark
Indicator Hold/Non-Hold
*
Non-hold
Hold
* When the alarm indication is set to non-hold, alarm ACK does not work.
Hold
Blinking red
icon
Blinking red
Blinking green
icon
icon
No icon
Alarm ACK
Alarm
Alarm Indication
Activated
Alarm mark, alarm type, measured
values : All red
Released
Alarm mark : Green color
Alarm type : Not displayed
Measured value : Blue color
Activated
Alarm mark : Blinks in red (Stops
blinking with alarm ACK)
Alarm type, measured value : Displayed
in red
Released
Alarm mark : Blinks in green (Stops
blinking with alarm ACK)
Alarm type : Not displayed
Measured value : Blue
Alarm
Alarm Indication
Activated
Alarm mark, measured values : All red
Released
Alarm mark : Green
Measured value : Blue
Activated
Alarm mark : Blinks in red (Stops
blinking with alarm ACK)
Measured value : Red
Released
Alarm mark : Blinks in green (Stops
blinking with alarm ACK)
Measured value : Blue
Activated
Released
Red icon
Alarm ACK
Measured value
Alarm mark