7.10 Setting the Scale Division, Bar Graph Base Position (Bar Graph), and Scale Position (Trend)
Explanation
Scale Specifications
Scale display position
On the trend display, the scale display position for the channels that are assigned to
groups can be displayed in 6 different positions. (See the figure below.)
• During trend vertical display
The scale display position is 6, 5, 4, 3, 2, and 1 from the top.
The scale is divided into 4 to 12 sections as shown in the figure above.
• If the scales for two or more channels are specified to the same position, the scale for
the channel that was assigned first to the group is displayed.
Example 1
If the scale display positions for channels 3, 2, 1, and 5 are all set to [1], the scale for
channel 3 is displayed at position 1.
Vacant positions in between scale assigned positions are void. The scales are displayed
close together from the display position 1.
Example 2
If the scale display position for channels 1, 2, 3, and 5 are set to positions 1, 3, 5, and
6, respectively, the scales for the channels are actually displayed at positions 1, 2, 3,
and 4, respectively.
• If the scale display position is set to [Off], the scale is not displayed.
Scale marks
The scale can be divided into 4 to 12 sections using the main scale marks. When the
scale is divided into 4 or 5 section, the area between the main scale marks is divided
further into 10 sections using small and medium marks. When the scale is divided into 6
to 12 sections, the area between the main scale marks is divided further into 5 sections
using small marks.
However, small marks are not displayed for the following cases :
• When the measurement/computation range resolution is smaller than the total number
of sections created by small marks.
• When zone display is used
• When partial expanded display is used
If the channels were assigned to a group in the following order :
[03.02.01.05]
If the channels were assigned to a group in the following order :
[01.02.03.05]
4 scale division sample
5 scale division sample
6 scale division sample
7 scale division sample
8 scale division sample
9 scale division sample
10 scale division sample
11 scale division sample
12 scale division sample