Omega RD-MV100 Series User Manual page 241

Hide thumbs Also See for RD-MV100 Series:
Table of Contents

Advertisement

Setting the absolute timer
2.
6.
7.
8.
9.
Confirming operation
To confirm the new settings, press the DISP/ENTER key.
To cancel the new settings, press the ESC key. A window appears for you to confirm the
cancellation. Select [Yes] using the arrow keys and press the DISP/ENTER key.
Explanation
Acquiring the data of all channels to the internal memory at each interval (setting
at the [Action] box)
If [Save data] is specified in the [Action] box, the instantaneous values of not only the
channels set to compute the TLOG computation, but also those of every channel can be
acquired to the internal memory at the specified interval (whether or not any channel is
set to compute the TLOG computation has no relevance).
The data along with the TLOG data are written to the area for writing the TLOG
computed data of the internal memory. When storing to the external storage medium,
the data are stored to the TLOG data file ("DTG" extension). For details related to the
data acquisition of data to the internal memory and the storage to the external storage
medium, see the explanation on the TLOG data in chapter 8.
Note
TLOG data are written to the internal memory at the interval determined by the timer when
the timer is specified (see section 11.8) (writing to the internal memory is carried out
regardless of the setting specified in the above-mentioned [Action] box).
Mode
Select [Absolute]. [Interval], [Reset], [Ref. time], and [Action] are displayed.
Interval
Set the time to the next timeout from 19 choices below.
1 min, 2 min, 3 min, 4 min, 5 min, 6 min, 10 min, 12 min, 15 min, 20 min, 30 min,
1 h, 2 h, 3 h, 4 h,6 h, 8 h, 12 h, 24 h.
Ref. time
Pressing the [Input] soft key displays a window used to enter the time. Enter the
time and press the DISP/ENTER key. For the procedures related to entering
numerical values, see "Entering Numbers" on page 3-21.
Reset
[On] : Resets the TLOG value when the timer expires.
[Off] : Does not reset the TLOG value when the timer expires.
Action
[DataSave] : Stores the instantaneous values of all measurement/computation
channels to the internal memory when the time expires.
[Off] :
Does not store the TLOG data.
11.9 Setting the Timer (Basic Setting Mode)