Omega RD-MV100 Series User Manual page 61

Hide thumbs Also See for RD-MV100 Series:
Table of Contents

Advertisement

Using the accessory DC power terminal connector
Follow the warnings below to avoid electric shock or damaging the instrument.
• Connect the power wires after checking that the power supply is turned off to
prevent electric shock.
• To prevent fire, use wires with cross sectional area of 0.3mm2(22AWG) or
more.
Connecting procedure
1. Check thst the power switch is turned off.
2. Connect the accessory DC power terminal connector (part No. A1105JC) to the power
wire and the DC100/DS400/DS600. Use a power wire with a cross sectional area of
2
0.3 mm
(22AWG) or more.
Install the clamp filter (standard accessory) near DC100 of power wire.
Rated supply voltage:
12 VDC
Operating supply voltage:10 to 18 VDC
Power consumption:
About 30 VA max.
2.7 Power Supply Wiring
WARNING
Soldering
-(0 VDC)
+(10 to 18 VDC)
2