Terumo Advanced Perfusion System 1 Operator's Manual page 82

Table of Contents

Advertisement

Central Control Monitor
To Calibrate O
Touch the Calibrate button.
Calibration takes approximately 2 minutes.
"Calibrating O
When the calibration is complete, the message "O
will display in the Messages subtab within the Aux tab and in the message area of the
CCM if there are no higher priority messages.
Pumps Subtab
The Pumps subtab allows the user to change the roller pump display units, tube size,
and pump response to a safety event. The Pumps subtab also allows the user to
change centrifugal pump display units, pump response to a safety event, and set the
minimum flow alarm. See the Roller Pumps or Centrifugal Control Unit chapter for
detailed information about connecting and operating pumps.
For Roller Pumps
Touch the Pump button until the correct name is displayed on the button. The roller
pump must be in Stop mode to change the display units and the tube size.
Pump Button
Tube Size Box
1/8 in.
6mm
S1:1
3/16 in.
8mm
S2:1
1/4 in.
10mm
S4:1
5/16 in.
12mm
S8:1
3/8 in.
Custom
7/16 in.
1/2 in.
Display Units Button
• L/min
• RPM
}
Primary
• L/min/m
2
Pump Only
• mL/kg/min
Only 3 events can be viewed
at one time. Use scroll arrows
to view other configured events
and pump response.
To Change Display Units:
Touch the Display Units button to toggle through the selections. Stop when the
desired selection is displayed. Changes are only retained for the current case.
To Change Tube Size:
1. Use the scroll arrows to view the list of available sizes.
2. Touch the correct tube size from the list. The tube size selection is highlighted
and the tube size will display above the selection box and on the roller pump local
display.
To enter a custom tube size:
1. Select Custom from the list.
2. Touch the Flow Constant button.
3. Use arrows to adjust value.
4. Touch button again to retain the value.
The system will retain this selection until a new selection is made from this
screen or from the pump local display.
Analyzer:
2
analyzer" is displayed during calibration.
2
Setup
Patient
Pump:
Arterial
Gas
Display Units:
Pumps
L/min
L/min
Occluders
Timers
Events
Time/Date
ART Pres Pres Alarm
ART Pres Pres Alert
ART Flow Back Flow
ART Pres Pres Alarm
Device Name
and Event
4-20
Analyzer Calibration Complete"
2
Type:
Large Roller Pump
Tube Size:
Custom
S4:1
Flow Constant
S8:1
FC144
Custom
Pump Response:
Stop : 360 mmHg
Reduce Speed : 300 mmHg
Stop
Stop
Response
Response Selection Button
16:07:51

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Advanced Perfusion System 1 and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

801763801764

Table of Contents

Save PDF