Central Control Monitor
Body Surface Area (BSA)
The BSA is calculated automatically when the height and weight data are entered.
Touch the Body Surface Area button to toggle through the selection for the formula
used to calculated the BSA. BSA can be calculated using the Dubois, Boyd, or
Pediatric formulas.
The Dubois formula is BSA = (Height)
The Boyd formula is BSA = 3.207 (Height)
The Pediatric formula is BSA = (Wt
Blood Volume
The Blood Volume is automatically calculated using the patient weight information
and is displayed in milliliters. If the patient weight is greater than or equal to 20 kg,
the formula used is 70 mL x Wt (kg). If patient weight is greater than or equal to 10 kg
and less than 20 kg, the formula is 75 mL x Wt (kg). If the patient weight is less than
10 kg, the formula is 80 mL x Wt (kg).
Gas Subtab
From the Gas subtab, the user selects the type of gas to blend with O
FiO
alarm and calibrates the oxygen analyzer. See the Electronic Gas Blender
2
chapter for detailed information about connecting and operating the gas blender.
Setup
Choices are:
Patient
Air
95/5
Gas
Pumps
Occluders
Timers
Time/Date
To Select O
Blend:
2
Touch the O
Blend button to toggle between Air and 95/5. The O
2
(95/5 or Air) will be displayed within the button.
To Set Low FiO
Alarm:
2
1. Touch the Low FiO
2
2. Use the arrows to change the alarm limit. The large arrows adjust the FiO
by increments of 0.5 and small arrows adjust the limit by increments of 0.1.
3. Touch the Low FiO
2
Note: The low FiO
Alarm is retained for the current case only and is reset each time
2
a Perfusion Screen is loaded.
With air selected as the O
range of 0.21 - 0.99. With 95-5 selected as the O
to "---" and has a range of 0.95 – 0.99. Asterisks are displayed and the alarm is
disabled if the alarm limit is set lower than the default value.
x (Weight)
.725
0.3
x (Weight x 1000)
10000
x Ht
) x 0.024265.
.5378
.3964
02 Blend:
Air
Low FiO2 Alarm:
Alarm button. The button changes to white.
Alarm button again to store the new value.
blend, the Low FiO
Alarm defaults to "---" and has a
2
2
4-19
x 0.007184.
.425
(0.7285 – 0.0188 Log (Weight x 1000))
, sets the low
2
16:07:51
O2 Analyzer:
Calibrate
Calibration Status:
Not Calibrated
blend selected
2
blend, the Low FiO
Alarm defaults
2
2
limit
2