Central Control Monitor

Pump Controls and Displays

Pump control boxes displayed on the lower third of the CCM can be used to start and
stop pumps and to monitor current pump speed or flow values. Each pump has a
corresponding slide bar located either on the right or left side of the Perfusion Screen
which can be used to adjust the pump speed on the CCM. Pumps can also be
controlled from their front panel controls.
Pump Control Box
All pump control boxes consist of three elements: a Start/Stop button, a speed/flow
display button, and a pump icon button. The only exception is a master/follower pump
control box (see Roller Pumps chapter for details of a master/follower pump control
box). The pump type, name, color, display units, and position are all defined in the
CCM Configuration area.
Centrifugal Pump Icon
Pump Name
Speed/Flow Display
Start/Stop Buttons
Start Button or Stop Button
The Start button activates and starts the pump at a speed of 0 RPM or a flow of
0.000 L/min. The Stop button stops the pump if it is running and also deactivates it.
If the pump is stopped and deactivated using the Stop button, it must be reactivated
using the Start button for the pump to be used again.
To start a pump:
1. Touch the Start button.
The button color changes to red and the label changes to Stop.
The pump speed/flow display button appears.
The pump slide bar appears on one side of the CCM (the primary pump slide bar
is always displayed).
To stop a pump:
1. Touch the Stop button.
The button color changes to blue and the label changes to Start.
If the pump is not the primary pump, the speed/flow button and the pump slide
bar disappear.
Pump Speed/Flow Display
The current value of the pump speed or flow rate is displayed in the units selected
during Configuration.
Touching the button on a primary pump will scroll through three possible display
selections: L/min, L/min/m
Note: If a centrifugal pump is not configured to display flow through the use of a
flowmeter, the button will display dashes.
For all other pumps, the pump speed/flow display button appears when the pump is
started and disappears when the pump is stopped. Touching the speed/flow display
button will open the slide bar for that pump and close any open slide bar in the
shared location.
Arterial
CPG
0.000
L/min
L/min
START
STOP
, or mL/kg/min.
2
4-13
Roller Pump Icon