Modules
Temperature Monitoring
The alarm will clear when the condition has been corrected.
ART Temp : OVER TEMPERATURE
Safety
Previous
ART Temp
Temps
43.0 C
Flow
Test Before Use
The performance of the temperature probes must be verified before each use.
Action
Cool probe below room temperature.
Warm probe above alarm level.
Temperature Monitoring Messages
Priority Condition
Alarm
Temperature has risen
above alarm limit.
Error
Temperature probe
has become
disconnected.
Temperature out of
range.
Limits ( C)
Alarm:
42.5
Results
Temperature decreases
Temperature icon flashes red. Temp alarm
displays on the CCM and the temp module LED
flashes red.
CCM Message
<Temperature name>:
OVER TEMPERATURE
<Temp channel name>:
Check Probe
<Temp channel name>:
Check Probe
7-23
16:07:51
Operator Response
Correct over temperature condition.
Alarm clears automatically.
Reconnect temperature probe.
Error message clears automatically.
Correct out of range condition.
Error message clears automatically.