Modules
Air Bubble Detection (ABD)
In the Event of an Air Alarm
The following actions should be taken if you receive an Air Alarm.
During Prime or Setup:
•
Ensure there is no air in the circuit.
•
Reset air alarm and restart pump(s), if required.
•
If alarm persists or repeats with no air present, check for the following setup
issues.
•
Check to ensure the tubing is properly placed inside the sensor and the latch
is fully and securely closed.
•
Check to ensure the air sensor cable is properly attached to both the sensor
and the ABD Module.
•
If no setup issues were found, consider backup air detection equipment.
•
Contact Terumo Customer Service.
During Cardiopulmonary Bypass:
•
If necessary, stop the arterial and/or cardioplegia pump.
CAUTION: Failure to stop the cardioplegia pump when stopping the arterial
pump may cause infusion of air into the circuit depending on the circuit setup.
•
Clamp the line, if appropriate.
•
Inspect for air and remove air, if needed.
•
Reset the alarm and restart the pump.
•
If alarm will not reset:
•
Disable the air sensor by turning it OFF on the CCM in the Air subtab within
the Safety tab. It will be necessary to use an alternate air monitoring method.
•
Use a backup Air Sensor and Cable, if possible.
•
Contact Terumo Customer Service after the case is completed.