Air Bubble Detection (ABD)

In the Event of an Air Alarm

The following actions should be taken if you receive an Air Alarm.
During Prime or Setup:
Ensure there is no air in the circuit.
Reset air alarm and restart pump(s), if required.
If alarm persists or repeats with no air present, check for the following setup
issues.
Check to ensure the tubing is properly placed inside the sensor and the latch
is fully and securely closed.
Check to ensure the air sensor cable is properly attached to both the sensor
and the ABD Module.
If no setup issues were found, consider backup air detection equipment.
Contact Terumo Customer Service.
During Cardiopulmonary Bypass:
If necessary, stop the arterial and/or cardioplegia pump.
CAUTION: Failure to stop the cardioplegia pump when stopping the arterial
pump may cause infusion of air into the circuit depending on the circuit setup.
Clamp the line, if appropriate.
Inspect for air and remove air, if needed.
Reset the alarm and restart the pump.
If alarm will not reset:
Disable the air sensor by turning it OFF on the CCM in the Air subtab within
the Safety tab. It will be necessary to use an alternate air monitoring method.
Use a backup Air Sensor and Cable, if possible.
Contact Terumo Customer Service after the case is completed.
7-6

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Advanced Perfusion System 1 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

This manual is also suitable for:

801763801764

Table of Contents

Save PDF