Terumo Advanced Perfusion System 1 Operator's Manual page 327

Table of Contents

Advertisement

Additional Information
Roller Pump
The fan filter, located on the bottom of the pump, should be replaced every 6 months
or more often if necessary. Use a 1/16 in. ball driver (provided with Accessory Kit) to
loosen the 2 screws and remove the panel. Remove the used filter and insert a new
air filter. Replace panel.
CCM
The CCM filter, located on the bottom of the CCM, should be replaced every 6
months or more often if necessary. Use a Phillips head screwdriver (provided with
Accessory Kit) to loosen the 2 screws and remove the panel. Remove the used filter
and insert a new filter. Replace the panel.
Lamp Bulb
When replacing the 24 VDC lamp bulb, hold the bulb with a cloth or tissue during
installation to prevent skin oils from shortening the life of the bulb.
Long Term Storage
Long term storage is defined as greater than one month without charging the backup
battery system by connecting the Advanced Perfusion System 1 to AC (Mains) power
and turning the System 1 on.
WARNING: Charge the batteries monthly. Storage of a System 1 without
charging the batteries monthly may damage the batteries and lead to overstated
battery life. This damage, caused by a gradual discharge of the batteries, may
cause loss of battery backup without warning.
If long term storage (beyond a month without a full charging cycle) of the System 1 is
desired, the batteries must be replaced before clinical use.
12-16