Roller Pumps
Priority
Condition
Status
Follower pump has
stopped in response to
master pump event.
Follower pump has
paused in response to
master pump event.
Master pump has
stopped in response to
follower pump event.
Master Pump has
paused in response to
follower pump event.
Pump pauses because
cover has been opened.
Pump direction has
been changed to
reverse.
Pulse mode has been
activated - base is
ramping down to target
value.
Manual button is
pressed when the
pump is in Pulse mode
or Servo mode.
Stop button is pressed
in Pulse mode or
Servo mode.
Pump is paused during
volume auto dose
delivery.
Pump is paused during
time auto dose delivery. <current time remaining>
Volume auto dose
delivery finished.
Time auto dose
delivery finished.
Pump stops during
volume auto dose
delivery.
Pump stops during
time auto dose
delivery.
Pump Display Message
STOPPED: MASTER
PAUSED: MASTER
STOPPED: FOLLOWER
PAUSED: FOLLOWER
PAUSED: COVER OPEN
PUMPING: REVERSE
RAMP BASELINE: XX%
MANUAL: MONITOR FLOW
MANUAL: MONITOR FLOW
PAUSED: <current volume> mL Message clears when pump
PAUSED:
DONE: <current volume> mL
DONE: 00:00
STOPPED: <current volume> mL Message clears when pump is
STOPPED:
<current time remaining>
5-31
Operator Response
Message should clear when
Master pump is restarted.
Message should clear when
master pump resumes.
Message should clear when
master pump is restarted.
Message should clear when
follower pump resumes.
Message should clear when cover
is closed.
Message should clear when default
direction is resumed. Default
direction is resumed when the
pump stops.
Monitor the speed change. The
base is reduced by 5% every 2
seconds. Message clears when
the target Base setting is reached.
Message should clear when speed
is changed or pump is stopped.
Message clears when pump is
started.
resumes or stops.
Message clears when pump
resumes or stops; time continues
to count down.
Message clears when auto dose
delivery is restarted.
Message clears when auto dose
delivery is restarted.
restarted; volume is reset to zero.
Message clears when pump is
restarted; time resumes counting
down.