Emergency Procedures
Roller Pump
Mode
Status
Start
Remains as set
Stop
Pause
Continuous
Servo Control
Remains as set
Pulse Mode
Remains as set
Master/Follower Remains as set
Centrifugal Control Unit
Mode/Alarm
Status
Start
Remains as set
Stop
Coast
Continuous
Servo Control
Remains as set
Pulse Mode
Remains as set
Min Flow
Remains as set
Alarm
Display
Flow rate or speed on pump
display
Setpoints and status displayed
on pump display
Pulse mode status displayed
on pump display
Master/Follower status, ratio, and Control from pump speed control
total flow displayed on pump
Display
Flow rate or speed on pump
display
Setpoints and status displayed
Pulse mode status displayed
on control unit display
Message on pump display of
centrifugal control unit and
connected pump(s)
10-13
Control
Complete control from knobs and
buttons on pump control panel.
Control from pump speed control
knob. Pump speed control knob can
be used to adjust setpoint.
Disable servo control using manual
button on pump control panel.
Cannot enable if disabled.
Disable Pulse mode control using
the manual button on pump
control panel.
Cannot enable if disabled.
knob on pump control panel.
Master pump speed control knob can
be used to change total flow rate.
Follower pump speed control knob
can be used to change the ratio or
percentage.
Control
Complete control from knobs and
buttons on centrifugal control unit.
Control from pump speed control
knob on centrifugal pump display
control unit. Pump speed control
knob can be used to adjust setpoint.
Disable servo control using manual
button on centrifugal control unit.
Cannot enable if disabled.
Disable Pulse mode control using
manual button on centrifugal control
unit.
Cannot enable if disabled.
No control, message clears when
condition clears.