Emergency Procedures
Air Bubble Detect Module
Condition
Status
On/Off
Remains as set
Air Detect
Functional
Alarm
Level Detect Module
Condition
Status
On/Off
Remains as set
Low Level
Functional
Alarm
Low Level Alert Functional
Pressure Monitoring
Condition
Status
Monitoring
Monitoring
capability is lost
High Pressure
Functional
Alarm
High Pressure
Functional
Alert
Temperature Monitoring
Condition
Status
Monitoring
Monitoring
capability is lost
Over Temp.
Functional
Alarm
Display
LED on module displays status
LED flashes red
Dual tone alarm sounds if
connected to a pump.
Message on pump display of
connected pump(s)
Display
LED on module displays status
LED flashes red
Dual tone alarm sounds if
connected to a pump.
Message on pump display of
connected pump(s)
LED flashes yellow
Single tone alarm sounds
if connected to a pump.
Message on pump display of
connected pump(s)
Display
LED on module displays status.
Setpoint displayed on pump
display if in Servo mode.
LED flashes red
Dual tone alarm sounds if
connected to a pump.
Message on pump display of
connected pump(s)
LED flashes yellow
Single tone alarm sounds
if connected to a pump.
Message on pump display of
connected pump(s)
Display
LED on module displays status
Connect sensors to portable
monitor to display temperatures.
LED flashes red
10-14
Control
Cannot enable if disabled.
Reset Air Bubble Detect module via
Select button on pump control panel
of connected pump when condition
is cleared.
Control
Cannot enable/disable.
No control, alarm clears when
condition clears.
No control, alert clears when
condition clears.
Control
Unable to calibrate.
No control, alarm clears when
condition clears.
Cannot change limits.
No control, alert clears when
condition clears.
Cannot change limits.
Control
None
No control, alarm clears when
condition clears.
Cannot change limits.