Centrifugal Control Unit
CAUTION: A pump may drop to a speed below 1500 RPM if the pump was
previously in pulsatile mode with an average speed lower than 1500 RPM when a
Coast response occurs.
Coast Lockout
When a safety event with a configured Coast response occurs, a centrifugal pump
will lock out manual adjustments to the pump speed for 2 seconds. This brief lockout
response guarantees that the safety response occurs.
The Coast lockout is signified in two ways during the brief lockout period:
The centrifugal pump slide bar on the CCM grays out and is disabled.
The Bar Graph on the Centrifugal Control Unit blinks.
User control is returned after the lockout period.
Coast Event
A Coast event is an event generated by a centrifugal control unit that can initiate
a response in other pumps and the occluder. The Coast event message is sent
when the centrifugal control unit goes to Coast speed (1500 RPM) as a response to
another event or when the user decreases the speed of the centrifugal control unit to
1550 RPMs. Devices configured to respond to a Coast event will do so regardless of
how the event was generated. For examples, see the illustration below.
User adjusts
pump speed
< 1550 RPM
Safety event
i.e. low
level alarm
Occluder responds
Pump goes
to Coast
goes to Close
Pump
Occluder responds
Response:
Coast
goes to Close
6-10
to Coast,
to Coast,