Teledyne T3DSO1000 Programming Manual page 66

Digital oscilloscopes
Hide thumbs Also See for T3DSO1000:
Table of Contents

Advertisement

DIGITAL
DESCRIPTION
COMMAND SYNTAX
QUERY SYNTAX
RESPONSE FORMAT
EXAMPLE
Note:
The table below shows the availability of command in each oscilloscope series.
Model
Digital Oscilloscope Series
The CUSTOM command sets the threshold
value by customer for the specified group of
channels. The threshold is used for triggering
purposes and for displaying the digital data as
high (above the threshold) or low (below the
threshold).
The CUSTOM? query returns the threshold
value set by customer for the specified group of
channels.
<group>:CUSTOM <value>
<group>:={H8,L8}
• H8 — D8-D15.
• L8 — D0-D7.
<value>:= volt value with unit.
Note:
• You need to add the volt unit(V/mV) to the
value. If there is no unit added, it defaults to
volts (V).
• The range of value varies from models. See
the data sheet for details.
• An out-of-range value will be adjusted to the
closest legal value.
<group>:CUSTOM?
<group>:CUSTOM <value>
For T3DSO1000 series, when the Digital
function is on, the following command sets the
threshold value of D8-D15 to 5V.
Command message:
L8:CUSTOM 5V
Valid?
CUSTOM | CUS
Command /Query
66