QUERY SYNTAX
RESPONSE FORMAT
EXAMPLE
Digital Oscilloscope Series
channels.
(first rising edge - last
falling edge)
Delay between two
LFR
channels.
(first falling edge - last
rising edge)
Delay between two
LFF
channels.
(first falling edge - last
falling edge)
Delay between two channels.
Skew
(edge – edge of the same
type)
<sourceA-sourceB>:MEASURE_DELY?
<type>
<sourceA-sourceB>:MEAD <type>,<value>
Model
T3DSO1000
others
The following steps show how to get the measured
value of phase between C2 and C4.
Step 1: Send the message to set the measurement
to Phase between C2 and C4, and then there
displays a phase measurement on the screen.
Command message:
MEAD PHA,C2-C4
Format of <value>
Numerical
value
in E-notation
with
SI
unit,
such
1.24E-04S.
Except
for PHA, it returns as
"44.65degree".
Numerical
value
in E-notation
with
SI
unit,
such
2.06E+01degree.
as
as
100