PASS/FAIL
DESCRIPTION
COMMAND SYNTAX
QUERY SYNTAX
RESPONSE FORMAT
EXAMPL
RELATED COMMANDS
Digital Oscilloscope Series
The PF_DISPLAY command displays
information in Pass/Fail test features.
The PF_DISPLAY? query returns whether the
message of Pass/Fail is displayed.
PF_DISPLAY <state>
<state>:={ON,OFF}
PF_DISPLAY?
PF_DISPLAY <state>
The following steps display the message of
Pass/Fail.
Step 1: Send command to set the Pass/Fail test
enable.
Command message:
PFEN ON
Step 2: Send command to display the message
of Pass/Fail.
Command message:
PFDS ON
PFEN
PF_DISPLAY | PFDS
Command /Query