ACQUIRE
DESCRIPTION
COMMAND SYNTAX
QUERY SYNTAX
RESPONSE FORMAT
EXAMPLE
Digital Oscilloscope Series
MEMORY_SIZE | MSIZ
The MEMORY_SIZE command sets the
maximum depth of memory.
The
MEMORY_SIZE? query
maximum depth of memory.
MEMORY_SIZE <size>
<size>:={7K,70K,700K,7M} for non-interleaved
mode. Non-interleaved means a single channel is
active per A/D converter. Most oscilloscopes
feature two channels per A/D converter. .
<size>:={14K,140K,1.4M,14M} for interleave
mode. Interleave mode means multiple active
channels per A/D converter.
Note:
Options of <size> vary from models. See the
data sheet or the acquire menu of the
oscilloscope for details.
MEMORY_SIZE?
MEMORY_SIZE <size>
The following command sets the maximum
depth of memory to 14M in interleave mode.
Command message:
MSIZ 14M
Command /Query
returns
the
31