MEASURE
DESCRIPTION
QUERY SYNTAX
RESPONSE FORMAT
QUERY SYNTAX
PARAMETER_VALUE? | PAVA?
The PARAMETER_VALUE query measures
and returns the specified measurement value
present on the selected waveform.
There are three uses for this command:
� Usage 1
<source>:PARAMETER_VALUE?
<parameter>
<source>:= {C1,C2,C3,C4}
<parameter>:={PKPK,MAX,MIN,AMPL,TOP,
BASE,CMEAN,MEAN,RMS,CRMS,OVSN,FP
RE,OVSP,RPRE,PER,FREQ,PWID,NWID,RIS
E,FALL,WID,DUTY,NDUTY,ALL}
See the table Description of Parameter for
details.
<source>:PARAMETER_VALUE
<parameter>,<value>
<value>:= Numerical value in E-notation with
SI unit.
� Usage 2
PARAMETER_VALUE? CUST<x>
<x>:= 1 to 5, and ALL
Digital Oscilloscope Series
Usage
Description
Usage 1
Specify the source and the
measurement.
See the command "MEAD?"
to get the measured value of
delay measurement.
Usage 2
Use "PAVA? CUST<x>" to
get customized.
Usage 3
Use " PAVA? STAT<x>" to
get statistics.
Custom
Description
Q u ery
105