ACQUIRE
DESCRIPTION
COMMAND SYNTAX
QUERY SYNTAX
RESPONSE FORMAT
EXAMPLE
Digital Oscilloscope Series
ACQUIRE_WAY | ACQW
The ACQUIRE_WAY command specifies the
acquisition mode.
The ACQUIRE_WAY? query returns the current
acquisition mode.
ACQUIRE_WAY <mode>[,<time>]
<mode>:={SAMPLING,PEAK_DETECT,AVE
RAGE,HIGH_RES}
<time>:={4,16,32,64,128,256,512,...}
• SAMPLING — sets the oscilloscope in the
normal mode.
• PEAK_DETECT — sets the oscilloscope in
the peak detect mode.
• AVERAGE — sets the oscilloscope in the
averaging mode.
• HIGH_RES — sets the oscilloscope in the
enhanced resolution mode (also known as
smoothing). This is essentially a digital boxcar
filter and is used to reduce noise at slower
sweep speeds.
Note:
• The [HIGH_RES] option is valid for T3DSO
models. See models on
• <time>:={4,16,32,64,128,256,512,...} when
<mode> = AVERAGE.
Options vary from models. See the data sheet or
the acquire menu of the oscilloscope.
ACQUIRE_WAY?
ACQUIRE_WAY <mode>[,<time>]
The following command sets the acquisition
mode to average mode and also sets the average
time to 16.
Command message: