Teledyne T3DSO1000 Programming Manual page 166

Digital oscilloscopes
Hide thumbs Also See for T3DSO1000:
Table of Contents

Advertisement

TRIGGER
DESCRIPTION
COMMAND SYNTAX
QUERY SYNTAX
RESPONSE FORMAT
EXAMPLE
Digital Oscilloscope Series
TRIG_LEVEL2 | TRLV2
The TRIG_LEVEL2 command sets the lower
trigger level voltage for the specified source.
Higher and lower trigger levels are used with
runt /slope triggers.
The TRIG_LEVEL2? query returns the lower
trigger level voltage for the specified source.
<trig_source>:TRIG_LEVEL2 <trig_level>
<trig_source>:= {C1,C2,C3,C4}
<trig_level>:= -4.5*DIV to 4.5*DIV.
Note:
• You need add the volt unit(V/mV) to the
trig_level. If there is no unit added, it defaults
to volts (V).
• An out-of-range value will be adjusted to
the closest legal value.
<trig_source>:TRIG_LEVEL2?
<trig_source>:TRIG_LEVEL2 <trig_level>
<trig_level>:= Numerical value in E-notation
with SI unit.
When the trigger type is slope, the following
steps set the high trigger level of Channel 2 to
3.5 V, and the low trigger level of Channel 2 to
800 mV.
Step 1: Send the command to set high trigger
level.
Command message:
C2:TRLV 3.5V
Step 2: Send the command to set low trigger
level.
Command message:
Command /Query
166

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

T3dso2000

Table of Contents