LOAD
QUERY SYNTAX
RESPONSE FORMAT
EXAMPLE
Note:
See the table Availability of WGEN Commands in Each Oscilloscope Series for details.
<load>
:={ HZ, 50}.
Digital Oscilloscope Series
WAVEGENERATOR? <parameter>
<parameter>:={OUTP,WVTP,FREQ,AMP
L,OFST,DCOFST,DUTY,SYMM,WIDTH,
STDEV,MEAN,LOAD,ALL}
WAVEGENERATOR
<parameter>,<value>
• For T3DSO2000 series, when the AWG
option is installed, the following command
set the type to square, amplitude to 2.5 V,
frequency to 10 kHz and duty to 45%.
Command message:
WGEN
TYPE,SQUARE,FREQ,10000Hz,AMPL,2.5
V,DUTY,45%
• For T3DSO2000 series, when the AWG
option is installed, the following command
set the type to noise, stdev to 0.2 V, mean to
1 V.
Command message:
WGEN
TYPE,NOISE,STDEV,0.2V,MEAN,1V
• For T3DSO2000 series, when the AWG
option is installed, the following command
set the output to off.
Command message:
WGEN OUPT,OFF