Teledyne T3DSO1000 Programming Manual page 41

Digital oscilloscopes
Hide thumbs Also See for T3DSO1000:
Table of Contents

Advertisement

CHANNEL
Command /Query
DESCRIPTION
COMMAND SYNTAX
QUERY SYNTAX
RESPONSE FORMAT
EXAMPLE
RELATED COMMANDS
Digital Oscilloscope Series
ATTENUATION | ATTN
The ATTENUATION command specifies the
probe attenuation factor for the selected channel.
The probe attenuation factor may be 0.1 to
10000.This command does not change the actual
input sensitivity of the oscilloscope. It changes
the reference constants for scaling the display
factors, for making automatic measurements,
and for setting trigger levels.
The ATTENUATION? query returns the current
probe attenuation factor for the selected channel.
<channel>:ATTENUATION <attenuation>
<channel>:={C1,C2,C3,C4}
<attenuation>:={0.1,0.2,0.5,1,2,5,10,20,50,100,2
00,500,1000,2000,5000,10000}
<channel>:ATTENUATION?
<channel>:ATTENUATION <attenuation>
The following command sets the attenuation
factor of Channel 1 to 100:1. To ensure the data
matches the true signal voltage values, the
physical probe attenuation must match the scope
attenuation values for that input channel.
Command message:
C1:ATTN 100
VDIV
OFST
41

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

T3dso2000

Table of Contents