TRIGGER
DESCRIPTION
COMMAND SYNTAX
(FOR ALL BUT TV)
Digital Oscilloscope Series
TRIG_SELECT | TRSE
The TRIG_SELECT command selects the
condition that will trigger the acquisition of
waveforms.
Depending on the trigger type, additional
parameters must be specified. These additional
parameters are grouped in pairs. The first in
the pair names the variable to be modified,
while the second gives the new value to be
assigned. Pairs may be given in any order and
restricted to those variables to be changed.
The
TRIG_SELECT?
current trigger condition.
Parameter
SLEW
Slope
GLIT
Glitch/
Pulse
INTV
Interval
DROP
Dropout
SR
Source
TI
Time
HT
H o l d
type/
Limit
range
HV
Hold
value/
Limit
value
TRIG_SELECT
<trig_type>,SR,<source>,HT,<hold_type>,H
V,<hold_value1>[,HV2,<hold_value2>]
Command /Query
query
returns
description
IL
Interval
larger
IS
Interval
smaller
I2
Interval
in range
I1
Interval
out
of
range
P
Pulse
PL
L
larger
PS
P u l s e
smaller
P2
Pulse in
range
P1
Pulse out
of range