CARESCAPE Monitor B650
Wrong character is displayed
when a barcode is read.
Barcode reader does not read a
multi-field barcode correctly.
(i.e., the information is not
populated correctly to the fields
in the Admit menu).
Speaker / audible alarm issues
Problem
Audible alarms do not work.

Incorrect system time

11.8.3
Problem
System time is incorrect when
patient monitor is not
connected to network.
System time is incorrect when
patient monitor is connected to
MC Network.
System time is incorrect when
patient monitor is connected to
S/5 network.
162
2081903-001
The keyboard locale is not
configured correctly.
The barcode reader's language
configuration is incorrect.
Barcode reader parser
configuration is incorrect.
The barcode reader is
incompatible with the parser
configuration (field lengths, field
types, delimeters, symbologies
etc.).
Possible cause
Audible alarms are turned off
(See Alarms Setup > Audible &
Visual.)
Alarm volume is low.
Speaker failure
Speaker cable loose or faulty.
Tone generator or audio
amplifier failure
Possible cause
CPU timekeeper battery empty
Time not configured properly
Network device time
synchronization error.
iCentral time incorrect.
Configure the keyboard locale correctly,
see "7.9".
Refer to the barcode reader manual.
Configure the barcode settings, see
"7.7.2".
Check the barcode information content
and compare it to the current parser
configuration.
Recommended action
Enable audible alarms.
Adjust alarm volume (Monitor Setup >
Sound Volumes).
Replace the speaker unit.
Check that the speaker cable is intact and
properly connected.
Replace CPU board .
Recommended action
Replace CPU timekeeper battery.
Configure date and time, see "7.3".
When adding a new device to the
CARESCAPE Network, the existing devices
on the CARESCAPE Network will
synchronize to the new device's time. To
prevent potential time synchronization
issues, you should set the new device's
time to be as close as possible to the time
(within one minute or less) used by the
existing GE devices on the CARESCAPE
Network.
Configure iCentral time master.