GE HEALTHCARE CARESCAPE Monitor B650 Technical Manual page 141

Hide thumbs Also See for CARESCAPE Monitor B650:
Table of Contents

Advertisement

Required Tools
E-PSM(P)(W) or E-(P)RE(S)TN module
A multiparameter patient simulator with invasive pressure adapter cable to GE invasive
pressure connector.
5 lead ECG trunk cable, IEC or AHA
5 leadwire set, IEC or AHA
Oscilloscope
Analog output cable (2000633-001)
NOTE: You can alternatively use the GE defib sync tester, 2040582-001, together with the
Analog output cable, 2000633-001 and a multiparameter simulator to perform this test. Follow
the instructions included with the tester. See analog output cable wire colors and related
signals from the table below.
Connecting cables
1.
Ensure that the module is connected to the patient monitor.
2.
Connect the 5-lead ECG trunk cable to the green ECG connector in the module.
3.
Connect the 5-leadwire sets to the trunk cable and to the simulator.
4.
Connect the invasive pressure adapter cable to the simulator and the other end to the
module.
5.
Connect the analog output cable to the defibrillation synchronization connector in the
patient monitor.
6.
Connect the open wires of the other end of the analog output cable to the oscilloscope
probe as described in each test. Use the figure and table below as a reference for making
the connections.
Defibrillation synchronization connector
Pin number
Signal
1
Digital defibrillator synchronization
marker out signal
2
Digital defibrillator synchronization
marker in signal
3
Common GND
4
Analog GND
5
Analog GND
6
IP analog output
7
ECG analog output
*) Brown wire is not connected.
Maintenance and checkout
Related wire color in the open
end of the Analog output cable
*)
Black
Green
Red
Blue
White
Yellow
Grey
133