4.3 Local access to Webmin with a service PC connected
to the MC connector
You can access Webmin locally by connecting an Ethernet crossover cable between the
service PC and the MC connector of the patient monitor.
NOTE: This connection method is available only if the monitor is equipped with the Standard
Interface Board that provides only the MC connector and the patient monitor is in a discharged
state.
NOTE: Ensure that there is no patient admitted, prior to disconnecting the monitor from a live
MC or S/5 Network.
NOTE: Normal patient monitoring is disabled in this connection method. You must exit the
maintenance mode to re-enable the normal patient monitoring mode. (Refer to
Webmin' on page 43.)
1.
Connect a service PC to the MC connector on the patient monitor using a crossover cable.
2.
Select Monitor Setup > Service calibrations.
3.
Log in with your biomed username and password and press Enter.
4.
Record the MC IP Address shown next to the Start Maintenance Mode –menu button:
MC IP Address: ______________________
MC Netmask: ______________________
5.
Select Start Maintenance Mode.
A screen saver with the text "Note: Monitor temporarily under maintenance" appears on
the screen.
6.
Configure the service PC's IP address and subnet mask to the same network segment
with the monitor's MC Network.
NOTE: For more information on how to configure the IP address, refer to the PC's
documentation.
7.
Launch a web browser on the service PC.
8.
In the Address field, type https://[MC IP address]:10000 and press Enter.
NOTE: [MC IP address] is the MC Network IP address of the patient monitor.