NOTE: If the IX IP address field is shown as 0.0.0.0., you need to configure the patient monitor's
IX Network address first. Access Webmin using the integrated browser and configure a static IP
address for the IX Network.
4.
Configure the service PC's IP address and subnet mask to the same network segment
with the patient monitor's IX Network setting.
NOTE: For more information on how to configure the IP address, refer to the PC's
documentation.
5.
Launch a web browser on the service PC.
6.
In the Address field, type https://[IX IP address]:10000 and press Enter.
NOTE: [IX IP address] is the IX Network IP address of the patient monitor.
The Login to Webmin dialog box displays.
7.
Continue to
4.5. Login to
Closing Webmin
1.
Restore the patient monitor's original IX Network configuration and service PC's network
settings if they were changed.
2.
Disconnect the crossover cable from the patient monitor and from the service PC.
3.
Reconnect the patient monitor back to the IX Network if applicable.
Webmin.
Using Webmin service interface
2081903-001
41