GE HEALTHCARE CARESCAPE Monitor B650 Technical Manual page 142

Hide thumbs Also See for CARESCAPE Monitor B650:
Table of Contents

Advertisement

CARESCAPE Monitor B650
Configuring the monitor
1.
Configure ECG:
a.
Select ECG1 to the screen with waveform and adequate priority.
b.
In the Parameter Setup > ECG > Setup, select ECG1 Lead: II.
2.
Configure invasive pressure:
a.
Select P1 to the screen with waveform and adequate priority.
b.
In the Parameter Setup > Invasive Pressures > P1, select:
Configuring the simulator
Refer to the simulator documentation for details on how to use and configure the simulator.
1.
Configure ECG:
2.
Only for modules with invasive pressure: configure invasive pressure:
Testing the synchronization connector
1.
Test the ECG analog output signal:
a.
Connect the oscilloscope probe to the ECG analog output signal (grey wire) and analog
ground (blue or white wire).
b.
Adjust the oscilloscope Time and Volts scales:
c.
Verify that the ECG analog output waveform shown on the oscilloscope screen
corresponds to the ECG1 waveform on the monitor screen.
2.
Test the Arterial BP output signal:
NOTE: Perform this arterial BP test only if an E-PSMP, an E-PSMPW, an E-PRETN or an E-
PRESTN module is in use with the patient monitor.
a.
Ensure that the simulator's InvBP output channel is configured to 0 mmHg static pressure
or atmosphere.
134
2081903-001
Label: P1.
Scale (mmHg): 0-200 mmHg.
Parameter Format: Sys/Dia (Mean).
ECG rhythm: a normal sinus rhythm.
Heart rate: 80 bpm,
Amplitude: 1 mV.
Sensitivity: 5 µV/V/mmHg.
InvBP output: "0 mmHg static pressure" or "atmosphere".
ECG analog output signal: Grey
Analog GND: Blue or White
Probe Type: x10
Time/Division: 200 mS
Volts/Division: 0.5V