GE HEALTHCARE CARESCAPE Monitor B650 Technical Manual page 207

Hide thumbs Also See for CARESCAPE Monitor B650:
Table of Contents

Advertisement

snaps
Reassemble in reverse order.
NOTE: Do not leave any dust or dirt between the LCD Display and the plastic display cover, or the picture
quality might deteriorate.
NOTE: Center align the user interface cable carefully to the connector in the CPU board and/or User Interface
Board.
NOTE: The keypad may be damaged when it is detached from the Front Unit Assembly. It is advised not to
reuse the old keypad, but to order a new one and replace it simultaneously with the Front Unit Assembly.
Disassembly and reassembly
2. Remove the four T10 screws in the top and
bottom corners behind the Display Unit.
[Torque 0.6 Nm]
3. Release the two snaps, one on each side of
the monitor. The snaps are located in the
middle behind the Right and Left Housing
Decorations.
4. Detach the WLAN card from the CPU if
installed.
NOTE: Be careful not to damage the antenna
cables.
5. Detach the Front Unit Assembly.
6. Disconnect the U/I cable from the User
Interface Board.
199
2081903-001