Restarting The Patient Monitor - GE HEALTHCARE CARESCAPE Monitor B650 Technical Manual

Hide thumbs Also See for CARESCAPE Monitor B650:
Table of Contents

Advertisement

CARESCAPE Monitor B650
WARNING
Control of this user's password is critical to ensure that Webmin on this
device is accessed only by trained and authorized personnel. Failure to limit
access of Webmin to trained and authorized personnel only may
compromise patient safety and/or system performance.
NOTE: Username and password are case sensitive. The allowed characters for "biomed" and
"clinical" passwords are: alpha [A-Z, a-z], numeric [0-9], and space. The password length shall
be between 8 and 16 characters.
NOTE: The user name "biomed" is common for the Webmin and Service Calibrations login
screens. A change to the "biomed" password will affect both service interfaces.
1.
Log in to Webmin.
2.
Select Configuration > Passwords.
3.
In the Edit User Password window, change the biomed or clinical user's password as
required.
4.
Select Save.
The change will take effect immediately.

7.14 Restarting the patient monitor

You can use the Restart module in Webmin to restart the patient monitor after making
configuration changes that require a manual restart to come into effect. For example after
changing network or language settings, or adding activation codes for licenses.
NOTE: Loss of monitoring - This function is enabled only when the patient monitor is in a
discharged state. Before restarting the patient monitor, verify that the patient is discharged
from the patient monitor.
1.
Log in to Webmin.
2.
Select Configuration > Restart.
3.
In the Monitor Restart window, select Restart.
The patient monitor will shutdown and restart automatically.
7.15 Setting up the remote service
7.15.1 Configuring the remote service
NOTE: Only available if the patient monitor is equipped with the Advanced Interface Board that
provides both the MC and the IX Network interfaces.
To configure the Insite with Exc remote service tool.
1.
Log in to Webmin.
2.
Select Configuration > Remote Service.
3.
In the Sub-Modules for Remote Service menu, select Configuration.
88